What does a STATIM do?
STATIM provides sterilization and dryness at speeds faster than conventional chambered autoclaves. Using only a small amount of water for each cycle, the STATIM heats and converts to steam in mere seconds. The innovative design of STATIM’s fully removable chamber allows for quick heating and cooling.
What is the difference between STATIM 2000 and 5000?
Larger than the 2000, the Statim 5000 has an additional sterilization cycle and a roomier cassette designed to process bulkier instruments and greater loads.
How long does a STATIM take to sterilize?
STATIM is easy to use. Instruments are sterilized at 135°C, with an overall cycle time of 6 minutes (STATIM 2000) or 9 minutes (STATIM 5000). Wrapped instruments are sterilized at 135°C, with a cycle time of 14 minutes (STATIM 2000) or 17.5 minutes (STATIM 5000). One-touch begins the sterilization process.
Is a STATIM a Class B?
The STATIM B G4+ Vacuum Autoclave is a compact EN13060 compliant Class B steriliser that offers you speed, high capacity, safety, and efficiency. Its innovative sterilisation technology incorporates both pre- sterilisation vacuum for greater steam penetration and post- sterilisation vacuum for improved drying.
How hot does a STATIM get?
Statim 5000 Wrapped Cycle depressurizes to 115°C. The unit will condition up and down between 115°C and 132°C six times. This depressurization is to remove air from the cassette and bags. The Solenoid then closes and the unit pressurizes to 132°C.
What is a dental autoclave?
Dental autoclaves are a commonly used method of cleaning and sterilizing dental equipment. An autoclave is a machine that is specifically designed to clean surfaces with extreme heat, steam, and a highly pressurized environment.
How hot does a dental autoclave get?
Autoclaves use saturated steam under pressure of approximately 15 pounds per square inch to achieve a chamber temperature of at least 250°F (121°C) for a prescribed time—usually 30–60 minutes.
How hot does the Statim 5000 GET?
Is a STATIM a pre Vacuum sterilizer?
More specifically, the STATIM® Units use Dynamic Air Removal Steam Sterilization, by the method of SFPP technology, and the BRAVOTM Units use Dynamic Air Removal Steam Sterilization, by the method of Pre-Vacuum.
Is a STATIM a pre vacuum sterilizer?
Why is my STATIM leaking?
If there is a small steam leak or water under the unit after installing the new seal try another cycle. It may take time to seat the new seal. If the leak persists, remove the seal and reinstall. If unit continues to leak after installing seal for the second time try a new cassette lid or tray.
Who makes STATIM?
SciCan
SciCan is proud to unveil the newest STATIM family member, the G4 series. The STATIM G4 series is the same renowned and trusted autoclave it has been for over 20 years but now boasts a new contemporary look, functionality and connectivity that is the first of its kind.